Xcalibur Diagnostics for the PDA Detector
Creating a Display Method to View the Light Intensity
46
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
To create a display method
1. Open the Surveyor PDA Plus Method page:
a. In the Home Page – Roadmap
view, click the
Instrument Setup
button.
The Instrument Setup view appears.
b. In the view bar, click the
Surveyor PDA
Plus
button.
The Surveyor PDA Plus Instrument Setup view appears (see
).
2. In the
Surveyor PDA Plus Method
page, enter the appropriate parameters:
a. In the Units area, select the
Diode/Intensity
option to set the X-axis of the spectrum
on the Display page to diode numbers.
b. In the Channels area, select the
Three Channels
option to make all three Channels
boxes available.
c.
In the Channels
boxes, type the numbers for the diodes that you want to monitor.
Figure 46.
Surveyor PDA Plus Method page with the selection of the Diodes/Intensity option
3. Save the display method:
a. In the menu bar, choose
Surveyor PDA Plus > Direct Control
.
shows the warning message that appears.
Diode/Intensity option