5
Routine and Preventive Maintenance
Maintaining the Probe Heater
78
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Installing the Probe Heater
Y
To install the probe heater
1. If you removed the O-ring from the underside of the probe heater base, reinstall it. See
Figure 61.
View of O-ring installed in the underside of the probe heater base
2. If you disconnected the grounding strap from the probe heater, reconnect it to the probe
heater. See
.
3. Rotate the black knobs of the probe heater, so that they face away from each other. See
.
4. Insert the tube portion of the probe heater into the probe mount.
5. Rotate the black knobs of the probe heater forward to their locking position. See
.
6. Plug the green connector and the white connector into their respective receptacles located
above the probe mount. See
.
Heater tube
O-ring, BS018
18.77 mm ID × 1.78 mm THK
CAUTION
Make sure that the grounding strap is connected to the probe heater before
you insert the probe heater into the probe mount.