Thermo Scientific Spin Tissue Processor Microm STP 120
34
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
Maintenance
Lubrication:
Every six months, raise the cover of the carrousel using the
€
key
to its uppermost position and lubricate it
exclusively with the
grease
supplied by the manufacturer.
Changing the filter:
Refer to instructions on changing the filter
Inspection
:
The spin tissue processor should be revised by a manufacturer
authorised technician at least once a year.
The processor should be cleaned every day just like any other
piece of laboratory equipment. Any spilt reagent should be
eliminated immediately as prolonged contact with the surfaces of
the unit could cause damage or stains.
CAUTION:
Before cleaning the processor, raise the cover using the
€
key
and then turn off the power supply at the Main ON/OFF switch.
DANGER:
When performing any maintenance or cleaning operations
always bear in mind that the reagents used may be highly
inflammable and that the paraffin baths may reach extremely
high temperatures.
n
Do not use reagents that contain acetone or xylol or any
abrasive powders.
n
Always clean the unit with domestic detergents without any
abrasive additives.
n
Use a plastic spatula to remove any remains of solid paraffin.
Do not use sharp objects that could scratch the surface.
n
The reagent vessels (not the paraffin baths) may be cleaned in
a conventional dishwater.
n
The paraffin baths may be disconnected from their bases for
easier cleaning.
CAUTION:
When replacing the paraffin baths, make sure that each one is
connected to the corresponding power plug.
n
During cleaning, make sure that no detergent liquid enters the
electrical connections and power plugs.
n
When cleaning the carrousel cover, use extreme care not to
force it in any way.
WARNING
:
Before using any cleaning or decontamination method or product
Cleaning
%
•
•
&