3
Description of the device
Workspace Components
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
3-21
Heating system
The workspace is heated by a surface heater system. The arrangement of the heating elements ensures
that condensate formation above the water reservoir is prevented.
The outer door of the unit and the perimeter of the door opening are heated as well. The heat radiated
onto the interior glass door/gas-tight screen prevents condensate formation.
The workspace of the device always remains visible, despite high humidity.
Rear panel openings
Figure 3-15.
Rear panel openings
The access port with insert (
Ø 42 mm / 1.66 inch) can be sealed with a plug
(1/
) and allows for feeding cables, tubes or additional sensors into the workspace of the
unit.
The pressure compensation opening (3/
) located directly beneath the access port in the rear
wall of the unit ensures pressure compensation between the workspace of the unit and the operating
room.
1
3
2
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...