12
Annex: Data communication
Error memory structure
12-8
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Note on
*3)
disinfection and auto-start run status:
Error memory structure
The error memory contains 22 error messages. A query is responded with 22 data sets with a colon as
separator and can be queried using the following commands:
Query:
?:2300:00::cc<CR>
Reading the first 11 error memory entries.
Query:
?:2301:00::cc<CR>
Reading the first 11 error memory entries.
These data sets consist of 11 byte and are encrypted in 21 ASCII characters prior to data transmission.
Example: The byte 0x23 is converted into the ASCII characters 0x32 (“2”) and 0x33 (“3”).
• Byte 1 consists of 1 characters,
• Bytes 2 - 11 consist of 2 characters.
Therefore, a response consists of 1+(10 x 2) = 21 data bytes plus separator.
A data set always delivers the date, the time, the faulty control loop, the device status, and the error
message.
Bit
Disinfection
auto-start
0x00
Steri-run not activated
auto-start not activated
0x01
Initialization
Initialization
0x02
Wait until open door time expires
Wait until open door time expires
0x03
Wait until door closes
Wait until door closes
0x04
Start
Start
0x05
Heating
Heating
0x06
Holding
Perform reverse voltage adjustment
0x07
Condensation
Waiting period 1
0x08
Cooling
Set tolerance range
0x09
Drying
Build up stable moisture
0x0A
Wait for enable
Perform reverse voltage adjustment
0x0B
Cancellation
Waiting period 2
0x0C
-
Determine offset
0x0D
-
Read out, check offset
0x0E
-
Release
0x0F
-
Cancellation
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...