11
Specifications
STERI-CYCLE i250
11-8
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
S
TERI-CYCLE
i250
Name
Unit
Value
CO
2
gas supply system
Gas purity
%
min. 99.5 or med. quality
Primary pressure
bar
min. 0.8 - max. 1 (11.6-14.5 psi)
Measurement and control range
Vol -%
0...20
Control deviation over time
Vol -%
± 0.1
CO
2
measuring cell
Accuracy at 37 °C and 5% CO
2
%CO
2
± 0.3
O
2
gas supply system
Gas purity
%
min. 99.5 or med. quality
Primary pressure
bar
min. 0.8 - max. 1 (11.6-14.5 psi)
Measurement and control range
VoI -%
1... 21 or 5...90
Control deviation over time
VoI -%
± 0.2
O
2
measuring cell
Accuracy at 37 °C and 21% O
2
%O
2
± 0.5 (option: 1.....21% O
2
)
± 2.0 (option: 5.....90% O
2
)
Electrical
Rated voltage
V
1/N/PE 230 V, AC (± 10%)
1/N/PE 220 V, AC (± 10%)
1/N/PE 120 V, AC (± 10%)
1/N/PE 100 V, AC (± 10%)
Rated frequency
Hz
50/60
Degree of protection (IEC 60529)
IP 20
Protection class
I
Overvoltage category (EN 61010)
II
Pollution degree (EN 61010)
2
Rated current
A
230 V:
Decontamination: 5.5
Incubation: 3.3
220 V:
Decontamination: 5.3
Incubation: 3.2
120 V:
Decontamination: 10.4
Incubation: 6.3
100 V:
Decontamination: 8.9
Incubation: 5.3
Circuit breaker
16 A
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...