9
Maintenance
Device fusing replacement
9-10
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Procedure for gas supply gas inlet filter:
1. Make sure that the gas supply is shut off.
2. Loosen the hose clamp (4/
).
3. Remove the gas hose (5/
) from the gas inlet filter sleeve.
Procedure for all gas inlet filters:
4. Remove the retainer (1/
).
5. Unscrew the gas inlet filter(2/
).
6. When installing the new gas inlet filter, make sure that the plastic thread is not canted. Install the
filter using caution.
7. Install the retainer.
Procedure for gas supply gas inlet filter:
8. Connect the gas hose to the gas inlet filter sleeve and secure it using the hose clamp. Check to see if
the gas hose is securely seated on the sleeve.
Figure 9-8.
Gas inlet filter installation
Device fusing replacement
The device’s fuses are not user-replaceable. If the unit fails due to an electrical fault, call Technical Field
Service.
1
2
3
5
4