7.3 Eliminating Leakage in the Solvent Rack
The leak sensor is installed inside the SRD and reports a leak when liquid collects in the
drip tray under the fluid connections. Locate the source for the leak, eliminate the cause,
and dry the leak sensor.
1.
Inspect the degasser inlets and degasser outlets, and the connected solvent supply lines
for signs of leaks. Tighten or replace leaking connections if necessary.
2.
Dry the components.
3.
With a cloth or tissue, absorb all liquid that has collected in the tray. Make sure that
you do not bend or damage the sensor.
Fig. 13: Drying the leak sensor
If the SRD is operated from the pump of the UltiMate 3000 system, the message remains
on the pump display for information purposes even after the source for the leak has been
eliminated. Select
Clear
on the navigation bar on the pump display to remove the
message.
4.
Allow the sensor to adjust to the ambient temperature for a few minutes.
5.
If no errors are reported, operation can be resumed.
Tip:
If the sensor is not dry, the Status LED remains red.
Содержание SRD-3200
Страница 2: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page II Operating Instructions ...
Страница 28: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 22 Operating Instructions ...
Страница 34: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 28 Operating Instructions ...
Страница 44: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 38 Operating Instructions ...
Страница 54: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 48 Operating Instructions ...
Страница 56: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 50 Operating Instructions ...
Страница 60: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 54 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 56 Operating Instructions ...