Maintenance and Care
Service of Thermo Fisher Scientific
Thermo Scientific
75006480 Diagnostik
TM
Rotor
5-5
For general radioactive decontamination, use a solution of equal parts of 70 % ethanol, 10 % SDS
and water.
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adapters and dispose of them or disinfect them.
6. Rinse the rotor first with ethanol and then with de-ionized water.
•
Adhere strictly to the given application times.
7. Be sure the decontamination solution can drain off the rotor.
8. Rinse the rotor and accessories thoroughly with water.
9. Dispose of the decontamination solution according to the applicable guidelines.
10. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 °C.
11. After cleaning, treat the entire surface of aluminum parts with corrosion protection oil
(70009824). Also treat the cavities with oil.
12. Treat the bolt of the swing out rotor with bolt grease (75003786).
Service of Thermo Fisher Scientific
Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an
authorized service technician. The service technician checks the following:
•
the electrical equipment;
•
the suitability of the set-up site;
•
the lid lock and the safety system;
•
the rotor;
•
the fixation of the rotor and the drive shaft.
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are
performed for free during the warranty period and afterwards for a charge.
This is only valid if the centrifuge has only been maintained by a Thermo Fisher Scientific service
technician.
CAUTION
Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Содержание Sorvall RC-4
Страница 4: ...C Contents ii 75006480 DiagnostikTM Rotor Thermo Scientific ...
Страница 7: ...Thermo Scientific 75006480 DiagnostikTM Rotor 1 1 1 Rotor Data Contents Technical Data on page 1 2 ...
Страница 11: ...Thermo Scientific 75006480 DiagnostikTM Rotor 2 1 2 Accessories Contents Accessory on page 2 2 ...
Страница 17: ...4 Rotor Loading Improper Loading Thermo Scientific 75006480 DiagnostikTM Rotor 4 3 Improper Loading ...
Страница 18: ...4 Rotor Loading Improper Loading 4 4 75006480 DiagnostikTM Rotor Thermo Scientific ...
Страница 24: ...5 Maintenance and Care Service of Thermo Fisher Scientific 5 6 75006480 DiagnostikTM Rotor Thermo Scientific ...
Страница 33: ......