1
Introduction and Description
Directives, Standards and Guidelines
1-4
Sorvall Legend XF / XFR
Thermo Scientific
Directives, Standards and Guidelines
Table 1-2.
Directives, Standards and Guidelines
Tension /
Frequency
Produced and inspected according to the following
standards and guidelines
230 V 50/60 Hz
nd
EN 61010-1, 2 Edition
EN 61010-2-020, 2
nd
Edition
EN 61326-1
EN ISO 9001
Functions and Features
The following table gives an overview of the important functional and perfor mance characteristics of
the Sorvall Legend XF / XFR.
Mains Supply
The following table contains an overview of the electrical connection data for the Sorvall Legend XF /
XFR. This data is to be taken into consideration when selecting the mains connection socket.
Table 1-3.
Functions and Features
Component / Function
Description / Features
Galvanized steel chassis with armoured plating
Structure / Housing
Stainless steel / varnished steel
Chamber
Induction drive without carbon brushes
Drive
Easy-to-clean keypad and display surface
Keys and display
Controls
Microprocessor-controlled
The most recent data is saved
Internal memory
Functions
RCF-selection, temperature control, pretemp with cooling equipped
devices
9 acceleration and 10 braking curves
Acceleration / braking profiles
Automatic
Rotor recognition
Electronic, contingent on rotor and speed
Imbalance recognition
Automatic lid closing and locking starting from an initial hold position
Lid lock
Table 1-4.
Electrical connection data of the Sorvall Legend XF / XFR
Cat.
Mains
voltage
Frequency
Rated
current
Power
consumption
Equipment
fuse
Building
fuse
refrigerated
230 V
±10 %
50 / 60 Hz
10.5 A
1950 W
15 AT
16 AT
75004561
ventilated
230 V
±10 %
50 / 60 Hz
11 A
1700 W
15 AT
16 AT
75004564