Troubleshooting & Service
The Transmitter
9-2
SGD-O User Guide
Thermo Fisher Scientific
This section provides troubleshooting procedures for the SGD-O
transmitter.
The instrument sends continuous readouts to the display. If the display is
blank, the readout is dim, or the readout displays in a yellow-green color,
increase the contrast by pressing the up arrow on the transmitter keypad.
If the display consists of dark rectangles, decrease the contrast by pressing
the down arrow.
If adjusting the contrast does not correct the display problem, first verify
the power supply at the source. If this is not the problem, remove power
from the transmitter. Open the enclosure and verify that the keypad cable
is properly installed on the CPU board.
If transmitter still does not display, the CPU board may be malfunctioning.
Refer to the SGD-O installation manual for instructions on replacing
boards in the transmitter.
If you suspect a problem with the DC power fuse, power down the unit
and measure the resistance with a multimeter. If the fuse is good, the
reading will be approximately 0 ohms.
You can also look at the fuse and determine it is bad if it looks brown.
Note
See
. If you are using a 12 Vdc board, the fuse is at F2.
Figure 9–1.
DC power fuses on the I/O relay board
The Transmitter
The Display
The DC Power Fuse
Содержание SGD-O
Страница 2: ......
Страница 3: ...SGD O Non Contacting Density Gauge For Fracturing Cementing User Guide P N 717910 Revision D...
Страница 4: ......
Страница 6: ...This page intentionally left blank...
Страница 8: ...This page intentionally left blank...
Страница 36: ...This page intentionally left blank...
Страница 66: ...This page intentionally left blank...
Страница 70: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...