Issue: 1.0
Technical Description - Circuitry
Thermo Fisher Scientific
SAM12 User Manual 7-5
The setting of high voltages and thresholds is done via the
serial interface.
The 1
st
,3
rd
and 4
th
counters for each channel form Cobalt
coincidence pulses which are available on OUT1 & 2
connectors.
The card has six connectors.
The scintillation detectors are connected via MHV connectors
on PMT1 & 2.
Coincident outputs are on BNC connectors OUT1 & 2.
X15 & X16
: These connectors provide power supplies to the
board and RS485 terminations:
X15 & X16 SIGNAL
1 EARTH
2 0V
3 RX-
4 TX-
5 Signal
Ground
6 TX+
7 +5V
8
RX+
9 +5V
Used in conjunction with PCB 42543-0223, this assembly
allows measuring the coincidence counting rate of up to 6
detectors. The number of coincidences can be determined
through setting the threshold no.1.
HV & Amplifier
Connections
Coincidence detector