4

6.2 Start Up

Refer to Section 6.1 and Figure 2 on page 3 as you complete the 
following procedures.

6.2.1 Turning the Power On

To start up your cryogenic freezer, complete the following steps:

1. Plug the freezer into the power outlet (refer to Section 4.2 on 

page 2).

2. Turn the key switch to the POWER ON position. The Power 

On LED and the Cabinet Temperature LED illuminate.

Note:

The alarm function is not active at this time. Refer to 
Section 5.2.3 for information about setting the alarms. In 

addition, the Setpoint Security will be deactivated in this 
key position.

3. The unit will start about 30 seconds later.

6.2.2 Setting the Cabinet Temperature

To set the cabinet temperature, complete the following steps:

1. Press and hold the Control Setpoint pad.

2. The Control setpoint LED lights and the Cabinet 

Temperature LED goes out.

3. Press and hold 

 to increase the temperature or 

 to 

decrease the temperature. The digital display scrolls through 
temperature settings.

4. Release both pads when the digital temperature display 

window shows the correct setpoint value.

Note:

If no keys are pressed within ten seconds, the 
temperature display reverts to the cabinet temperature.

6.2.3 Setting the Alarms

To set the cold alarm, complete the following steps:

1. Confirm that the key switch is at the Power On position 

(Setpoint Security disabled) 

2. Press and hold the Cold Alarm Setpoint pad. The LED next 

to this pad lights. The temperature display shows the Cold 

Alarm value.

3. Press and hold 

 or 

 to adjust the Cold Alarm Setpoint.

4. Release both pads when the digital temperature display 

window shows the correct setpoint value.

To set the warm alarm, complete the following steps:

1. Confirm that the key switch is at the Power On position 

(Setpoint Security disabled).

2. Press and hold the Warm Alarm Setpoint pad. The LED next 

to this pad lights. The temperature display shows the Warm 

Alarm value.

3. Press and hold 

 or 

 to adjust the Warm Alarm Setpoint.

4. Release both pads when the digital temperature display 

window shows the correct setpoint value. 

When the cabinet temperature drops below the Warm Alarm 
setting, turn the key switch to ALARM ON. The freezer is ready 
to operate.

Note:

If a power failure lasting over 30 seconds occurs, the 

POWER ON light goes out and the POWER FAILURE 
light and digital temperature display flash 
simultaneously.

WARNING!

You must turn the three-position key 

switch to the ALARM ON/SETPOINT SECURITY 
position to activate the alarm and place it in security 
operation.

6.2.4 Alarm Reset & Status Lights 

The alarm reset feature ensures user acknowledgment of the 
occurrence of certain alarm conditions. This provides greater 
monitoring power by alerting the user if an alarm condition has 

occurred during periods when the freezer must be left 
unattended. The alarm conditions incorporating this feature are:

Temperature Failure-Warm Alarm

Temperature Failure-Cold Alarm

Voltage Low

Power Failure

Life-Guard Alert

Extreme Ambient Alert 

During any of these alarm conditions, the corresponding 
indicator will flash quickly, approximately 90 times per minute. 

If the alarm condition then disappears, the flash rate will 
decrease to approximately 15 times per minute. The indicator 
will then remain flashing at this rate, unless the condition 

reoccurs, until the Alarm Reset button is pressed. Once this has 
occurred, the indicator will no longer be lit.

In the case of TEMPERATURE FAILURE-WARM ALARM and 

TEMPERATURE FAILURE-COLD ALARM, if these 
conditions have occurred but are now within the given limits, the 
two indicators will alternately flash (180° out of phase). In 

addition, for these two temperatures failures, the highest 
temperature and lowest temperature, respectively, during the 
error condition will be saved. Prior to the temperature failure 

being reset, the temperature extreme may be viewed by pressing 
the Cabinet Temperature pad in conjunction with either the 
Warm Alarm Setpoint pad or the Cold Alarm Setpoint pad.

Note:

Once the Alarm Reset button is depressed, the extreme 
temperature value is reset to the current cabinet 
temperature.

6.2.5 Alarm Test 

To test the cold and warm alarms, complete the following steps:

1. Press and hold the Alarm Test pad. The digital temperature 

display indicates rising temperature. When the temperature 
reaches the warm alarm value, the alarm sounds.

Note:

The temperature of the refrigerated space does not 
change during this procedure. Only the sensor is heated.

2. Release the Alarm Test pad, the sensor and display return to 

cabinet temperature in a few minutes.

Содержание Revco Ultima II

Страница 1: ...Installation and Operation Manual Thermo Scientific Revco Ultima II Cryogenic Storage Freezers...

Страница 2: ...maflex is a registered trademark of Armacell International GmBH All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights rese...

Страница 3: ...3 6 2 Start Up 4 6 2 1 Turning the Power On 4 6 2 2 Setting the Cabinet Temperature 4 6 2 3 Setting the Alarms 4 6 2 4 Alarm Reset Status Lights 4 6 2 5 Alarm Test 4 6 3 Backup System Optional 5 6 3 1...

Страница 4: ...hermo Scientific Revco Cryogenic Storage Freezers Installation and Operation ii 8 Chart Recorders 8 8 1 Set Up and Operation 8 8 2 Power Supply 8 8 3 Changing Chart Paper 8 8 4 Calibration Adjustment...

Страница 5: ...ation system is designed to maintain cryogenic temperatures with safety only when the freezer is used for storage These models may be used in an ambient environment up to 25 C 77 F WARNING This unit i...

Страница 6: ...ilation no air exchange a steady refrigerant leak Under these conditions the minimum free volume requirements would be at least 86 m3 or 3006 ft3 5 2 Wiring CAUTION Connect the equipment to the correc...

Страница 7: ...s pad to set the cabinet temperature refer to Section 6 2 2 on page 4 12 Cabinet Temperature Indicator This indicator lights when the temperature display window is showing the cabinet temperature refe...

Страница 8: ...tch to ALARM ON The freezer is ready to operate Note If a power failure lasting over 30 seconds occurs the POWER ON light goes out and the POWER FAILURE light and digital temperature display flash sim...

Страница 9: ...the cylinders in a handtruck or cart with secure chain ties for the cylinder After cylinders are connected to the equipment securely attach them with chains to a solid stationary object such as a bui...

Страница 10: ...s accurate to 1 C The backup system can run for a minimum of eight hours on battery power WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 3 4 System...

Страница 11: ...r standing in the bottom of the cabinet 6 If there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaning wax 7 Close the...

Страница 12: ...pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the left side of...

Страница 13: ...entifizieren die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli...

Страница 14: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Страница 15: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 39125H51 Rev D...

Отзывы: