Thermo Scientific Revco Slimline Value Freezers

Installation and Operation

1

1 Introduction

Revco

®

 Slimline Value -86ºC upright freezers feature the use of 

vacuum insulation technology to minimize wall thickness and 
provide extra storage space. Because of the thin wall 

construction, SI models can hold more inventory than 
conventional models that take up the same lab space.

Standard features include:

High storage capacity — up to 15 inventory racks

Separately mounted inner doors with adjustable hinges for 

each door

A pressure equalization port to combat the vacuum created 

after door openings

SI Series freezers use the Legaci

 refrigeration system. a 

combination of advanced technologies designed specifically to 

provide optimal reliability in low-temperature applications.

Legaci systems include:

Copeland compressors custom-designed to minimize bearing 

wear and operate efficiently under heavy work loads

Dupont Suva

® 

refrigerants selected to provide maximum 

cooling capacity under low-temperature conditions

A field-proven cascade refrigeration system design.

2 Safety Precautions

In this manual and on labels attached to this product, the words 
WARNING and CAUTION mean the following:

WARNING: a potentially hazardous situation which, if not 
avoided, could result in serious injury or death.

CAUTION: a potentially hazardous situation which, if not 
avoided, may result in minor or moderate injury or damage to 
the equipment.

Before installing, using or maintaining this product, please be sure 
to read this manual and product warning labels carefully. Failure to 
follow these instructions may cause this product to malfunction, 
which could result in injury or damage.

Below are important safety precautions that apply to this product:

Use this product only in the way described in the product 
literature and in this manual. Before using it, verify that this 
product is suitable for its intended use. 

Do not modify system components, especially the controller. 
Use OEM exact replacement equipment or parts. Before use, 
confirm that the product has not been altered in any way.

Your unit must be properly grounded in conformity with 
national and local electrical codes. Never connect the unit to 
overloaded power sources.

Disconnect the unit from all power sources before cleaning, 
troubleshooting, or performing other maintenance on the 
product or its controls.

3 Pre-Installation

3.1 Unpacking

At delivery, examine the exterior for physical damage while the 
carrier’s representative is present. If exterior damage is 

present, carefully unpack and inspect the unit and all 
accessories for damage.

If there is no exterior damage, unpack and inspect the 

equipment within five days of delivery. If you find any damage, 
keep the packing materials and immediately report the damage 
to the carrier. 

Do not return goods without written 

authorization

. When submitting a claim for shipping damage, 

request that the carrier inspect the shipping container and 
equipment.

4 General Recommendations

This refrigeration system is designed to maintain ultra-low 
temperatures with safety in a +32

°

C (90

°

F) ambient 

environment, 

only

 when the freezer is used for storage.

WARNING!

This unit is not a “rapid-freeze” device. 

Freezing large quantities of liquid, or high-water 
content items, will temporarily increase the chamber 
temperature and will cause the compressors to 
operate for a prolonged time period. 

Avoid opening the door for extended time periods since chamber 
temperature air will escape rapidly. Room air, which is higher in 

humidity, replacing chamber air may cause frost to develop in the 
chamber more rapidly.

4.1 Initial Loading

After reading and completing the Safety Considerations, 

Pre-Installation, Installation, and Operation sections of this 
manual, turn the Key Switch to the POWER ON position and 
adjust the temperature setpoint if necessary.

The setpoint should be no warmer than -50ºC. 

Allow the freezer 

to operate at the desired temperature for a minimum of 12 
hours before loading.

Load the freezer one shelf at a time, beginning with the top shelf. 
After loading each shelf, allow the freezer to recover to the 
desired setpoint before loading the next shelf. Repeat this process 

until the freezer is fully loaded.

CAUTION!

Failure to follow these procedures or 

overloading the unit may cause undue stress on the 
compressors or jeopardize user product safety.

Содержание Revco Slimline

Страница 1: ...Installation and Operation Manual Thermo Scientific Revco Slimline Value Ultra Low Temperature SI Freezers ...

Страница 2: ... registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Страница 3: ...Operation 4 7 1 Control Panel Features 4 7 2 Start Up 4 7 2 1 Turning the Power On 4 7 2 2 Setting the Cabinet Temperature 4 7 2 3 Alarm 4 8 Maintenance and Troubleshooting 5 8 1 Condenser Maintenance 5 8 1 1 Cleaning the Condenser 5 8 1 2 Cleaning the Condenser Filter 5 8 2 Gasket Maintenance 5 8 3 Defrost Procedures 5 8 4 Alarm Battery Maintenance 5 9 Chart Recorders 6 9 1 Set Up and Operation 6...

Страница 4: ...y Your unit must be properly grounded in conformity with national and local electrical codes Never connect the unit to overloaded power sources Disconnect the unit from all power sources before cleaning troubleshooting or performing other maintenance on the product or its controls 3 Pre Installation 3 1 Unpacking At delivery examine the exterior for physical damage while the carrier s representati...

Страница 5: ...afety and trouble free operation this unit must be properly grounded before it is used Failure to ground the equipment may cause personal injury or damage to the equipment Always conform to the National Electrical Code and local codes Do not connect the unit to overloaded power lines Your freezer is equipped with a service cord and plug designed to connect it to a power outlet which delivers the c...

Страница 6: ...within 30 seconds after door openings To minimize warm air entry the port automatically closes as soon as pressures are equalized The pressure equalization port is located on the right side of the storage chamber Although the port is designed to self defrost excessive frost accumulation in the bottom of the chamber could eventually restrict air flow Therefore you should periodically inspect the po...

Страница 7: ... 1 Insert a small screwdriver into the slotted screw labeled Control Setpoint and simultaneously press and hold the Setpoint Display button The temperature display changes to read the existing setpoint value 2 Turn the setpoint screw clockwise for a colder setting and counterclockwise for a warmer setting until the desired setpoint shows in the digital temperature display 3 Release the Setpoint Di...

Страница 8: ... 4 on page 2 for leveling information Keep the lid and door gaskets clean and frost free by wiping gently with a soft cloth 8 3 Defrosting the Freezer Defrost the freezer once or year or whenever the ice buildup exceeds 3 8 To defrost complete the following steps 1 Remove all products and place in another cabinet 2 Turn off the freezer 3 Open the outer door and all inner doors 4 Let the freezer st...

Страница 9: ...ery The LED indicator glows continuously when main power is functioning and the battery is charged When the battery is low the LED flashes to indicate that the battery needs to be changed 9 3 Changing Chart Paper To change the chart paper complete the following steps 1 Locate the pressure sensitive buttons at the front upper left of the recorder panel 2 Press and hold the Change Chart button 3 for...

Страница 10: ... is required Chart Paper Surge Suppressor Alarm Systems Remote Alarm Monitoring Description Catalog No For All Models Fiberboard boxes and grid dividers are packaged in lots of one dozen Fiberboard box 5 1 4 in square x 2 in high one dozen 5954 Fiberboard box 5 1 4 in square x 3 in high one dozen 5956 Grid divider fiberboard 100 cell 7 16 in holds 100 12 mm vials one dozen 5958 Grid divider fiberb...

Страница 11: ...en die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fisher Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europe...

Страница 12: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Страница 13: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 310655H01 Rev C ...

Отзывы: