background image

Installation and Operation Manual

Thermo Scientific Revco

®

 Slimline Value

Ultra-Low Temperature SI Freezers

Содержание Revco Slimline

Страница 1: ...Installation and Operation Manual Thermo Scientific Revco Slimline Value Ultra Low Temperature SI Freezers ...

Страница 2: ... registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Страница 3: ...Operation 4 7 1 Control Panel Features 4 7 2 Start Up 4 7 2 1 Turning the Power On 4 7 2 2 Setting the Cabinet Temperature 4 7 2 3 Alarm 4 8 Maintenance and Troubleshooting 5 8 1 Condenser Maintenance 5 8 1 1 Cleaning the Condenser 5 8 1 2 Cleaning the Condenser Filter 5 8 2 Gasket Maintenance 5 8 3 Defrost Procedures 5 8 4 Alarm Battery Maintenance 5 9 Chart Recorders 6 9 1 Set Up and Operation 6...

Страница 4: ...y Your unit must be properly grounded in conformity with national and local electrical codes Never connect the unit to overloaded power sources Disconnect the unit from all power sources before cleaning troubleshooting or performing other maintenance on the product or its controls 3 Pre Installation 3 1 Unpacking At delivery examine the exterior for physical damage while the carrier s representati...

Страница 5: ...afety and trouble free operation this unit must be properly grounded before it is used Failure to ground the equipment may cause personal injury or damage to the equipment Always conform to the National Electrical Code and local codes Do not connect the unit to overloaded power lines Your freezer is equipped with a service cord and plug designed to connect it to a power outlet which delivers the c...

Страница 6: ...within 30 seconds after door openings To minimize warm air entry the port automatically closes as soon as pressures are equalized The pressure equalization port is located on the right side of the storage chamber Although the port is designed to self defrost excessive frost accumulation in the bottom of the chamber could eventually restrict air flow Therefore you should periodically inspect the po...

Страница 7: ... 1 Insert a small screwdriver into the slotted screw labeled Control Setpoint and simultaneously press and hold the Setpoint Display button The temperature display changes to read the existing setpoint value 2 Turn the setpoint screw clockwise for a colder setting and counterclockwise for a warmer setting until the desired setpoint shows in the digital temperature display 3 Release the Setpoint Di...

Страница 8: ... 4 on page 2 for leveling information Keep the lid and door gaskets clean and frost free by wiping gently with a soft cloth 8 3 Defrosting the Freezer Defrost the freezer once or year or whenever the ice buildup exceeds 3 8 To defrost complete the following steps 1 Remove all products and place in another cabinet 2 Turn off the freezer 3 Open the outer door and all inner doors 4 Let the freezer st...

Страница 9: ...ery The LED indicator glows continuously when main power is functioning and the battery is charged When the battery is low the LED flashes to indicate that the battery needs to be changed 9 3 Changing Chart Paper To change the chart paper complete the following steps 1 Locate the pressure sensitive buttons at the front upper left of the recorder panel 2 Press and hold the Change Chart button 3 for...

Страница 10: ... is required Chart Paper Surge Suppressor Alarm Systems Remote Alarm Monitoring Description Catalog No For All Models Fiberboard boxes and grid dividers are packaged in lots of one dozen Fiberboard box 5 1 4 in square x 2 in high one dozen 5954 Fiberboard box 5 1 4 in square x 3 in high one dozen 5956 Grid divider fiberboard 100 cell 7 16 in holds 100 12 mm vials one dozen 5958 Grid divider fiberb...

Страница 11: ...en die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fisher Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europe...

Страница 12: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Страница 13: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 310655H01 Rev C ...