background image

Select 

Pre-Programmed Methods

 in the Main Menu.

 

Select the number of gels and gel size (mini or midi) for transfer:
a. 

1 mini-gel

 b. 

2 mini-gels or 1 midi-gel

c. 

3 mini-gels

 d. 

4 mini-gels or 2 midi-gels

Select the appropriate program to run: 
a. 

Low MW (< 25kDa)

b. 

Mixed Range MW (25-150kDa)

c. 

High MW (>150kDa)

d. 

Std Semi-Dry

e. 

1.5mm thick gels or if unknown

Note

: For fast-blotting programs (program a, b, c and e) use  

Pierce 1-Step Transfer Buffer. Transfer time may be increased to  
12 minutes for high molecular-weight proteins (> 200kDa). Do not 
use the Std Semi-Dry transfer program (program d) with Pierce 
1-Step Transfer Buffer.

Select the 

Start

 button to begin transfer. 

Pierce 1-Step Transfer Buffer is a highly concentrated salt solution. 

Thoroughly

 wash the anode and cathode after each use by rinsing 

the unassembled cassette under hot water while removing any salt 
residue with a gloved hand. Rinse with deionized water and stand 
parts in a rack to dry. For more thorough cleaning, immerse the un-
assembled cassette top (cathode) and bottom (anode) in hot water 
and use a gloved hand or clean sponge to remove salt residue. 
Rinse with deionized water and stand parts in a rack to dry. 

Note: Failure to keep cassette 

top and bottom clean can result 

in sticking of movable parts and 

lead to poor transfer efficiency.

IMPORTANT

  Review and implement guidelines for proper set-up 

before installation. For detailed instructions refer to the manual at  

www.thermoscientific.com/pierce-power-system.

Using the supplied power cord, connect the Pierce Power Station to an 
electrical outlet.

Press the Power Button (located at the rear) to turn ON the Pierce 
Power Station. 

IMPORTANT:

 For first use, refer to “

Using the Pierce Power 

Station for First Time

” section. If you currently use a Pierce Power 

Station for staining and are adding blotting  functionality, refer to  

Using the Pierce Power Blot Cassette for First Time

” section. 

Upgraded Pierce G2 Fast Blotter/Stainer Control Units do not require 
any special activation in order to recognize the Power Stain Cassette  
or the Power Blot Cassette.  

For each gel, use four sheets of ~0.83mm thick Western blotting filter 
paper and one sheet of nitrocellulose or PVDF membrane cut to the 
same size.

Equilibrate filter paper and membrane in Thermo Scientific™ Pierce™ 
1-Step Transfer Buffer for at least 5 minutes. Use sufficient buffer to 
cover filter paper and membrane [~50mL per mini-sized (7cm x 8.4cm) 
sandwich and ~100mL per midi-sized (8cm x 13.5cm) sandwich].

Note: 

PVDF membrane must be wetted with methanol or ethanol 

before equilibration in Pierce 1-Step Transfer Buffer.

After electrophoresis, remove gel(s) from cassette(s) and briefly place 
into tray containing deionized water or transfer buffer. This will ensure 
even wetting, facilitate proper gel placement and improve gel contact 
with membrane.

Assemble and center sandwich(es) on the anode (see figure below) 
to ensure even pressure. When transferring two gels allow for 10mm 
spaces between sandwiches. Use a blot roller to remove any trapped 
air bubbles.

Lock top of the cassette (cathode) into place and slide the assembled 
cassette into the Pierce Power Station.

Select 

Begin Blotting

.

1

2

3

4

5

6

7

8

10

Quick Start Guide

Thermo Scientific Pierce Electrophoretic Blotting  

11

9

12

Thermo Scientific

 Pierce

 Power Blotter  (# 22834), Thermo Scientific

 Pierce

 Power System (# 22830),

or Thermo Scientific

 Pierce

 G2 Fast Blotter (# 62291 upgraded with # 62287), and Thermo Scientific

 Pierce

 Power Blot  Cassette (# 22835)

13

Cathode (-)
2 sheets of pre-wet filter paper (combined 

thickness not to exceed 1.8mm)
Gel
Membrane
2 sheets of pre-wet filter paper (combined 

thickness not to exceed 1.8mm)
Anode (+)

Choose Program

1 mini-gel

Back

Low MW

6 V        1.3 A        

5:00

Mid-Range MW

11 V       1.3 A        

7:00

High MW

18 V      1.3 A      

10:00