background image

Cond

1413

DI

1

3

5

7

9

2

4

6

8

10

Conductivity Automatic Calibration 
with One Standard

1. Press 

the 

power

 key to turn on the meter.  

See 1

.  

2  In the measurement mode, press the 

line select

 key 

until the   icon points to the middle line and press 
the 

up arrow

 key until the 

μS/cm 

or

 mS/cm

 icon is 

shown.  

See 2

.

3.  Select the Thermo Scientifi c Orion conductivity 

standard (100 μS/cm, 1413 μS/cm or 12.9 mS/cm) that 
has the closest conductivity to the expected sample 
value.  

See 3

4.  In the setup mode, enter the nominal cell constant 

value for the conductivity probe.  See the 

Nominal 

Cell Constant Selection

 section.  

See 4

.  The 

meter will use the nominal cell constant value 
to recognize the conductivity standard during an 
automatic calibration.  

5.  Rinse the conductivity probe (and stirrer probe, if in 

use) with deionized water and blot dry with a lint-free 
tissue.  

See 5

.

6.  Insert the conductivity probe into the conductivity 

standard and gently stir.  

See 6

.  If the stirrer probe 

is in use, press the 

stir

 key to start and stop stirring 

(benchtop models only).

7. Press 

the 

calibrate

 key.  

See 7

.  

8.  The meter will show the manual calibration display 

for about fi ve seconds.  

See 8

.  Do not press any keys.   

9.  After about fi ve seconds, the meter will proceed to 

the direct and automatic calibration display.  Wait for 
the 

μS/cm

 or 

mS/cm

 icon to stop fl ashing and the    

icon to start fl ashing.  The meter should display the 
conductivity standard value at 25 ˚C.  

See 9

.

10. Press the 

measure

 key to save and end the 

calibration.  

See 10

.  The calculated cell constant 

will be displayed and the meter will proceed to the 
measurement mode.

Nominal Cell Constant Selection

1.  In the measurement mode, press the 

setup

 key.

2. Press 

the 

up arrow

 key until 

COnd

 is displayed on 

the top line.

3. Press 

the 

line select

 key to move the   icon to the 

middle line.  Press the 

up arrow

 key until 

CELL

 is 

displayed on the middle line.

4. Press 

the 

line select

 key to move the   icon to 

the bottom line.  To enter the nominal cell constant, 
press the 

digit

 key until the fi rst digit to be changed 

is fl ashing, press the 

up/down arrow

 keys to 

change the value of the fl ashing digit and continue 
to change the digits until the meter displays the 
correct value.  Once the value is set, press the 

digit

 

key until the decimal point is in the correct location.  
0.475 cm

-1

 is the default setting.

5. Press 

the 

line select

 key to move the   icon to the 

top line and press the 

measure

 key.

Reference Temperature Selection

1.  In the measurement mode, press the 

setup

 key.

2. Press 

the 

up arrow

 key until 

C0nd

 is displayed on 

the top line. 

3. Press 

the 

line select

 key to move the   icon to the 

middle line.  Press the 

up arrow

 key until 

trEF

 is 

displayed on the middle line. 

4. Press 

the 

line select

 key to move the   icon to 

the bottom line.  Press the 

up arrow

 key to select 

5

 °C, 

10

 °C, 

15

 °C, 

20

 °C or 

25

 °C as the reference 

temperature.  25 °C is the default setting.

5.   Press the 

line select

 key to move the   icon to the 

top line and press the 

measure

 key.

Automatic Datalog Feature

1.  In the measurement mode, press the 

setup

 key.

2. Press 

the 

up arrow

 key until 

LOg

 is displayed on 

the top line.

3. Press 

the 

line select

 key to move the   icon to the 

middle line.  Press the 

up arrow

 key until 

AUt0

 is 

displayed on the middle line.

4. Press 

the 

line select

 key to move the   icon to the 

bottom line.  Press the 

up arrow

 key to select 

OFF

 

or 

On

.  OFF is the default setting.

5. Press 

the 

line select 

key to move the   icon to the 

top line and press the 

measure

 key.

Orion Star

 Plus Conductivity Meter Quick Start Guide

Conductivity Measurements

AUTO-READ™ is the default measurement mode.  
Continuous or timed measurement modes can be 
selected in the meter setup menu.  

Turn on the automatic datalog feature to send 
measurements to the meter datalog or connect the 
meter to a printer or computer.

1.  Rinse the conductivity probe (and stirrer probe, if 

in use) with distilled water, blot dry with a lint-
free tissue and insert into the sample.

2.  If the meter is in the AUTO-READ mode, press 

the 

measure

 key to start a reading.  Once the 

reading is stable, the 

AR

 icon will stop fl ashing 

and the display will freeze.  Press the 

measure

 

key to take a new measurement.  If the stirrer 
probe is in use, it will start stirring when the 

measure

 key is pressed and stop stirring 

when the 

AR

 icon stops fl ashing (benchtop 

meters only). 

 

If the meter is in the continuous or timed mode, 
the meter will immediately start taking readings 
and continuously update the display.  Once the 
reading is stable, the 

μS/cm 

or

 mS/cm

 icon will 

stop fl ashing.  If the stirrer probe is in use, press 
the 

stir

 key to start and stop stirring (benchtop 

meters only).

3.  Remove the conductivity probe (and stirrer probe, 

if in use) from the sample, rinse with distilled 
water, blot dry with a lint-free tissue, insert into 
the next sample and repeat step 2 or see the 
probe user guide for recommended storage.

Printing Data

If the automatic datalog feature is on, the meter will 
log a measurement every time the 

measure

 key is 

pressed (AUTO-READ and continuous modes) or the 
meter will log the measurement at the preset time 
interval (timed mode).

1.  Connect the meter to a printer or computer and 

verify the baud rate and output settings in the 
meter setup menu. 

2.  In the measurement mode, press the 

log

 

view

 key.

3. Press 

the 

up arrow

 key to select 

SEnd

 to print 

the datalog or 

CALS

 to print the calibration log. 

4. Press 

the 

log view

 key to send the selected data 

to the printer or computer.  

English

Отзывы: