Thermo Scientific Orion 960 Titrator Addendum
For the stability handling during the electrode calibration in the
autosampler mode, electrode potentials will be read based on the following
criteria:
1. Stir for 10 seconds.
2. The 960 uses stability calculation based on the current 960 algorithm.
3. A reading will be taken when electrode reaches stability criteria of less
than 5.0 mV/min.
4. Before moving to the next method 960 will print the electrode channel,
Eo, and slope.
Note:
After 3 minutes, the 960 will take a reading regardless of the stability
criteria. A new slope and Eo will substitute the 940 slope and Eo.
Stability Handling