5
Maintenance
Instrument Diagnostics
126
NanoDrop One
C
with NanoDropQC Software User Guide
Thermo Scientific
2. Enter each lot-specific target absorbance value from the label on the PV-1 ampoule in its
associated entry box and then tap
Done
.
3. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
4. Pipette 1 µL DI H
2
O onto the lower pedestal, lower the arm and tap
Blank
.
5. Lift the arm and clean both pedestals with a new wipe.
6. Pipette 1 µL PV-1 solution onto the lower pedestal and start the sample measurement:
• If
is On, lower arm
• If Auto-Measure is off, lower arm and tap
Measure
or from the PC Control software,
select
Run
.
After the measurement, the software displays the results. Here is an example of the
performance verification result screen.
Note
Vigorously shake the ampoule of PV-1 solution, allow the liquid to collect in
the bottom portion of the ampoule and then follow standard practices to open it.
Tap an entry box to display