Measure OD600
Thermo Scientific
NanoDrop One User Guide
135
Setting
Available Options
Description
Absorbance correction
Absorbance value
between 0 and 300 A
User-defined absorbance correction. Enter absorbance correction
for displayed spectrum. This can be useful, for example, to
correct baseline offset caused by any difference between the
media solution used to blank the instrument and media used to
suspend the cell culture sample, and because scattered light
generally produces an offset.
Absorbance correction value is subtracted from absorbance values
at all wavelengths in sample spectrum. (All displayed absorbance
values are corrected values.)
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...