WARNING
Do not attempt to open lid until rotor has completely stopped!
Avoid severe personal injury or property damage from moving parts.
Only use rotor compatible with the centrifuge.
WARNING
Do not use damaged or cracked rotors!
Rotor installation and removal
To remove rotor:
•
With one hand, grasp the rotor. With the other hand, unscrew center knob counter
clockwise until it disengages from the motor shaft.
•
Pull up on the center knob firmly. The rotor should easily lift off of the motor drive.
To install the rotor:
•
Check to make sure the bands on the rotor are in the slots.
•
Set rotor onto the motor drive and push down until seated.
•
With one hand, grasp the rotor. With the other hand, screw the center knob clockwise
until it engages the motor shaft. Keep turning the knob until it tightens.
•
Do not overtighten the knob.