Maintenance
Disposal of materials
80
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Figure 5–32.
Unfastening the priming vessel
2.
Rinse or wash the priming vessel with water or detergent when
necessary.
3.
Fasten the dried priming vessel by sliding it back over the snap lock
The priming vessel is made of
blue
polypropylene (PP) and can be
autoclaved. The autoclaving conditions of the priming vessel are:
1 bar pressure at 121
°
C for 20 minutes.
Follow laboratory and country-specific procedures for the disposal of
biohazardous or radioactive waste. Refer to local regulations for the
disposal of infectious material.
Warning
The samples can be potentially infectious. Dispose of all used
plates, strips, disposable gloves, syringes, disposable tips, and so on as
biohazardous waste.
s
If you have spilled infectious agents, carry out the decontamination
procedure.
Warning
The decontamination procedure should be performed by
authorized and trained personnel in a well-ventilated room wearing
disposable gloves, protective glasses and clothing.
s
Decontamination should be performed in accordance with normal
laboratory procedures. Any decontamination instructions provided with
the reagents used should be followed.
Disposal of
materials
Decontamination
procedure
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...