Maintenance
Regular and preventive maintenance
72
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
10.
Prime with distilled and filtered laboratory-grade water.
11.
Reconnect the valve tube ends to the dispensing valves
Figure 5–26.
Reconnecting a dispensing valve tube end
12.
Finally prime once again with distilled and filtered laboratory-grade
water.
The Multidrop Combi nL liquid path components can be disinfected
with isopropanol, 10% bleach (for example, sodium hypochlorite), or
70% ethanol by running the disinfectant through the liquid path.
To obtain optimal performance and a maximum useful life for the
dispensing valves, it is important that the recommended cleaning
maintenance instructions are followed.
If the dispensing valve is leaking or clogged, refer to “How to clean a
leaking dispensing valve” on page 73 and/or “How to clean a clogged
dispensing valve” on page 75.
Routine maintenance of the dispensing valves covers the sections
described below.
The basic maintenance procedure should be performed regularly and on
a daily basis to ensure proper dispensing valve operation:
1.
Filter all reagents, detergents, solutions and water before use to
remove solid particles.
Disinfecting the liquid
path
Routine
maintenance of the
dispensing valves
(tips)
Daily maintenance
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...