Power Switch:
The power switch is a combination ON/OFF and mode
switch to enable TOUCH operation. Move the power switch
to either the “FULL” position or the “TOUCH” position. In
the “FULL” position, the mixing head will begin oscillating
automatically. In the “TOUCH” position, the mixing head
will oscillate when it is depressed.
Speed Control:
The speed control is a nonlinear type control. The oscillat-
ing speed will increase at a non-uniform rate as the control
is turned clockwise.
To operate Maxi-Mix, move power switch to either the
“FULL” position or the “TOUCH” position. Turn speed con-
trol to the desired setting. Touching the vessel to the disc
will initiate the vortex action. Mixing action will intensify as
greater pressure is exerted against the mixing surface. To
stop the mixing action, remove vessel from disc or move
the power switch to the “OFF” position. (See Maintenance
and Servicing for installation of the single cup mixing head
and the mixing surface disc).
Operation
Warning
Do not turn the mixer on unless the
speed control is turned to minimum;
failure to do so may cause acces-
sories and vessels to be thrown off
the mixer.
Note
There is no absolute limit on the
maximum size of vessel which can
be used; much depends on tech-
nique, the shape of the vessel, the
viscosity of the contents, and the
amount of liquid. (Vessels require
closure tops to prevent contents
from being spilled out during mix-
ing). As a general rule, do not vortex
vessels larger than 250 ml and vol-
umes greater than 125 ml.
Note
Unit may need some loading to slow
the motor when going from fast to
slow.
Note
If the equipment is used in a manner
not specified by the manufacturer,
the protection provided by the equip-
ment may be impared.