background image

7

Power Switch:

The power switch is a combination ON/OFF and mode

switch to enable TOUCH operation. Move the power switch

to either the “FULL” position or the “TOUCH” position. In

the “FULL” position, the mixing head will begin oscillating

automatically. In the “TOUCH” position, the mixing head

will oscillate when it is depressed.

Speed Control:

The speed control is a nonlinear type control. The oscillat-

ing speed will increase at a non-uniform rate as the control

is turned clockwise.

To operate Maxi-Mix, move power switch to either the

“FULL” position or the “TOUCH” position. Turn speed con-

trol to the desired setting. Touching the vessel to the disc

will initiate the vortex action. Mixing action will intensify as

greater pressure is exerted against the mixing surface. To

stop the mixing action, remove vessel from disc or move

the power switch to the “OFF” position. (See Maintenance

and Servicing for installation of the single cup mixing head

and the mixing surface disc).

Operation

Warning

Do not turn the mixer on unless the

speed control is turned to minimum;

failure to do so may cause acces-

sories and vessels to be thrown off

the mixer.

Note

There is no absolute limit on the

maximum size of vessel which can

be used; much depends on tech-

nique, the shape of the vessel, the

viscosity of the contents, and the

amount of liquid. (Vessels require

closure tops to prevent contents

from being spilled out during mix-

ing). As a general rule, do not vortex

vessels larger than 250 ml and vol-

umes greater than 125 ml.

Note

Unit may need some loading to slow

the motor when going from fast to

slow.

Note

If the equipment is used in a manner

not specified by the manufacturer,

the protection provided by the equip-

ment may be impared.

Содержание M37610-33

Страница 1: ...Maxi Mix II Vortex Mixer Operation Manual and Parts List LT1254X1 9 19 08 MODEL NUMBERS M37615 M37614 M37610 33 M37610 33CN...

Страница 2: ...6 Specifications 6 Operation 7 Power Switch 7 Speed Control 7 Maintenance and Servicing 8 To Replace the Power Switch 8 To Replace Rheostat Speed Control 8 To Replace Microswitch 9 To Replace Motor 10...

Страница 3: ...Laboratory Use Part I General Requirements EN 61010 2 051Part II Particular Requirements for Laboratory Equipment for Mixing and Stirring per the provisions of the Low Voltage Directive 73 23 EEC as...

Страница 4: ...rly grounded electrical outlet of cor rect voltage and current handling capacity 2 Disconnect from power supply before servicing To avoid personal injury 1 Do not use in the presence of flammable or c...

Страница 5: ...switch and 3 an adjustable speed control See Figure 1 Principles of Operation The mixer incorporates a permanently lubricated motor to drive the mixing platform A three position power switch enables...

Страница 6: ...Inches Model No Net Ship H L Top Diameter Volts Amps Hz M37615 6 7 6 7 3 5 120 45 50 60 M37610 33 6 7 6 7 3 5 230 20 50 M37610 33CN 6 7 6 7 3 5 230 20 50 M37614 6 7 6 7 3 5 100 56 50 60 Warning Use a...

Страница 7: ...face To stop the mixing action remove vessel from disc or move the power switch to the OFF position See Maintenance and Servicing for installation of the single cup mixing head and the mixing surface...

Страница 8: ...Step d g Reinstall plastic housing and secure with four screws h While holding the mixing platform to keep it from turning secure the platform to the counter weight assembly If the mixing surface is i...

Страница 9: ...ll plastic housing and secure with four screws k While holding the mixing platform to keep it from turning secure the platform to the counterweight assembly If the mixing surface is in good condition...

Страница 10: ...g surface is in good condition reapply it to the mixing platform k Reconnect mixer to power supply To Replace Motor a Disconnect mixer from power supply b Peel off the mixing surface disc Hold the mix...

Страница 11: ...turning and remove the platform from the counterweight assembly c Turn mixer upside down and remove four screws securing plastic housing d Turn mixer upright and remove housing e Remove defective damp...

Страница 12: ...Disconnect mixer from power supply b Peel off mixing surface disc and while holding the mixing platform to keep from turning remove mixing platform c While holding the rubber mixing cup to keep from...

Страница 13: ...Speed control at minimum setting and too much pres sure on disc Defective speed control Defective motor Check mixer connections to power supply Replace microswitch Turn speed control to higher setting...

Страница 14: ...4 14 Key No Part No Description Description Key No Part No This is not found on exploded view Replacement Parts List Ordering Procedures Please refer to the Specification Plate for the complete model...

Страница 15: ...15 Exploded View Exploded View...

Страница 16: ...use inconsistent with written operating instructions for the product mishandling contamination overheating modification or alteration of the product by any customer or third party or iii use of repla...

Отзывы: