Tuning with Your Analyte in LC/ESI/MS Mode
Setting Up to Introduce Sample by Syringe Pump into Solvent Flow from an LC
72
LTQ XL Getting Started
Thermo Electron Corporation
Figure 30.
ESI/MS plumbing connections for the LC Tee union
8. If you have not already done so, connect the PEEK safety sleeve and
fused-silica sample tube from the grounding union to the ESI probe
sample inlet as described in the section,
Installing a New Fused-Silica
Sample Tube and PEEK Safety Sleeve
in the
Ion Max API Source
Hardware Manual
.
If you have installed the stainless steel needle in the ESI probe, connect
the PEEK safety sleeve and fused-silica capillary tube from the
grounding union to the ESI probe sample inlet as described in the
section,
Installing a New Stainless Steel Needle in the ESI Probe
and
Installing a New Fused-Silica Sample Tube and PEEK Safety Sleeve
in the
Ion Max API Source Hardware Manual
.
9. Connect an appropriate length of PEEK tubing (transfer line from the
divert/inject valve) from the divert/inject valve to the free end of the
(black) LC Tee union, as follows.
a. Connect a length of PEEK tubing with a (stainless steel) nut and a
(stainless steel) ferrule to port 3 of the divert/inject valve.
See
Grounding Union
(P/N 00101-18182)
LC Tee Union
(P/N 00101-18204)
From LC Union
Fingertight
Fittings
(P/N 00101-18081)
Stainless Steel
Ferrule
Infusion Line
Fused-Silica Capillary
(P/N 00106-10504)
PEEK Tubing
(P/N 00301-22912)
From Divert /
Inject Valve
Ferrules
(P/N 00101-18196)
Ferrules
(P/N 00101-18120)
Stainless
Steel Nut
Fingertight
Fitting
(P/N 00101-18195)
PEEK Tubing
(P/N 00301-22912)