Acquiring APCI Sample Data Using the Tune Plus Window
Acquiring APCI Data in the SIM Scan Mode
134
LTQ XL Getting Started
Thermo Electron Corporation
Acquiring APCI Data
in the SIM Scan Mode
Acquire a file of reserpine data in the selected ion monitoring (SIM)
scan mode.
1. If you have not already done so, in Tune Plus, click
On/Standby
to take
the MS detector out of Standby mode and turn it On. The MS detector
begins scanning, and the LTQ XL MS detector applies high voltage to
the corona needle and shows a real-time display in the Spectrum view.
2. Ensure that the Centroid data type is selected. (The picture on the
button should be the same as that shown here.)
3. Ensure that the scan parameters are defined to acquire SIM data for
reserpine (or your analyte of interest), as follows:
a. Click the
Define Scan
button to open the Define Scan dialog box.
b. Verify that the values in your dialog box are the same as those in
Figure 8-3. Then, click
OK
.
Figure 65.
Define Scan dialog box, showing typical settings for acquiring data in the SIM scan mode
4. Turn on your LC pump, and specify an appropriate flow rate of
400
μ
L/min, for example. Verify that your system is free of leaks.
Note
The LTQ XL MS detector automatically saves the acquired data
on the hard disk.
On
Standby