Acquiring ESI Sample Data Using the Tune Plus Window
Acquiring MS Data in the SIM Scan
94
LTQ XL Getting Started
Thermo Electron Corporation
Acquiring MS Data in
the SIM Scan
Acquire a file of reserpine data using the selected ion monitoring (SIM)
scan from the Tune Plus window
1. On the Control/Scan Mode toolbar, click
On/Standby
to take the MS
detector out of Standby mode and turn it On. The MS detector begins
scanning, the LTQ XL MS detector applies high voltage to the ESI
probe, and a real-time display shows in the Spectrum view.
2. Verify that the Centroid data type is selected. (The picture on the
button should be the same as that shown here.)
3. Verify that the scan parameters are defined to acquire SIM data for
reserpine (or your analyte of interest), as follows:
a. Click the Define Scan button to open the Define Scan dialog box.
b. Compare the values in your dialog box to those in
click
OK
.
Figure 45.
Define Scan dialog box, showing typical settings for acquiring reserpine data in the SIM scan type
4. Turn on the LC pump and specify your flow rate of 400
μ
L/min. Ensure
that your system is free of leaks.
Note
The LTQ XL MS detector automatically saves the data you acquire
on your hard disk.