Standardization & Calibration
Perform Calibration
8.
When the time remaining is 0 seconds, click
Next
two times. The
software will display that the cycle has been completed. Click
Write
Changes
.
Figure 5–5.
Standardization complete
Perform
Calibration
erform
Calibration
Caution
If you do not calibrate the gauge, some tanks might overflow
without reaching the tank full level. This can occur if the radiation
measured at the detector when the tank is full (process level at the top of
the detector) is greater than the assumed background radiation. If your
application requires accurate level measurements, perform a calibration.
▲
Note
It is a good idea to calibrate at the tank full level first, then at several
intermediate levels, especially where the tank or gauge geometry changes.
▲
Follow these steps for each calibration point:
1.
If necessary, set up the gauge.
2.
Standardize according to
.
5-4
LevelPRO with
F
OUNDATION
™ fieldbus
Application Guide
Thermo Fisher Scientific