INSTALLATION

SHIPPING CARTON:

This should be inspected upon delivery. When received, carefully 

examine for any shipping damage before unpacking. If damage is discovered, 

the delivering carrier should both  specify and sign for the damage on your copy 

of the delivery receipt.

Open the carton carefully making certain that all parts are accounted for 

before packaging materials are discarded—after unpacking, if damage is found,

promptly report it to the carrier and request a damage inspection promptly.

IMPORTANT: Failure to request an inspection of damage within a few 

days after receipt of shipment absolves the carrier from any liability for damage: 

you must call for a damage inspection promptly.

LOCATION:

 Place the unit on a level table or bench where it will be operated, away 

from drafts and wide variations in ambient temperature, near a power supply 

source that matches the unit nameplate requirements. Allow clearance around 

the unit for free air convection, accessory attachment and user convenience.

The incubator door opens from the left side.

THERMOMETER INSTALLATION:

The thermometer holder doubles as a vent fixture on the top of the unit. 

Place the grommet in the hole at the top of the thermometer holder. Push the 

thermometer, supplied with the unit, through the grommet until 3-inches of the 

thermometer (including the bulb) extends through the holder. Push the holder 

into the vent on top of the cabinet until about 1/4-inch of the holder is exposed. 

NOTE: 

IT MAY BE NECESSARY TO SQUEEZE THE 3 PRONGS OF THE 

THERMOMETER HOLDER TO INSTALL IT MORE EASILY.

The thermometer bulb should extend about 2-inches into the chamber.

Turn the thermometer in its holder for easy reading.

THERMOMETER

THERMOMETER HOLDER GROMMET

THERMOMETER HOLDER

THERMOMETER VENT

(TOP OF INCUBATOR)

INSTALLATION:

 (Con’t)

SHELF INSTALLATION:

100 series comes with 1 fixed shelf.  120 series units are supplied with 

two aluminum shelves: one is fixed to the bottom of the chamber, the other is 

Содержание Lab-Line 100

Страница 1: ...e Compact Low Cost Incubators OPERATION AND REPAIR MANUAL AND PARTS LIST Model Numbers 100 150 Lab Line 100JPN 150JPN 100IDE 152 100 5JPN 101 105 106 120 120JPN 120 5JPN 121 125 126 Additional comment...

Страница 2: ...REV J LAB LINE COMPACT LOW COST INCUBATORS MODEL NO 100 100IDE 100JPN 100 5JPN 101 105 106 120 120JPN 120 5JPN 121 125 126 150 150JPN 152 1999 North 15th Ave Melrose Park IL 60160 1491 USA PHONE 563...

Страница 3: ...3 TABLE OF CONTENTS SECTION TITLE 1 Introduction 2 Description 3 Specifications 4 Installation 5 Operation 6 Maintenance 7 Replacement Parts Warranty...

Страница 4: ...his unit becomes familiar with the Operation Section of this manual Be sure to fill out the Warranty Registration Card and mail it in to Lab Line Instruments within seven 7 days after receiving the un...

Страница 5: ...elp to distribute heat evenly throughout the chamber Powder coated durable heavy gauge steel exteriors make the units easy to clean and maintain A bimetallic thermostat controls temperature Models 105...

Страница 6: ...JPN 152 13 7 16 W x 13 11 32 D x 12 11 16 H 36 x 34 x 32 cm EXTERIOR DIMENSIONS 100 IDE 100JPN 100 5JPN 101 105 106 14 3 8 W x 13 11 32 D x 11 5 16 H 37 x 34 x 30 cm 120 120JPN 120 5JPN 121 125 126 14...

Страница 7: ...nit nameplate requirements Allow clearance around the unit for free air convection accessory attachment and user convenience The incubator door opens from the left side THERMOMETER INSTALLATION The th...

Страница 8: ...Because of the variety of plug configurations in use worldwide for 240 VAC power the unit is furnished with the plug removed The user must install a plug to conform with local code and configuration r...

Страница 9: ...ght hand side of incubator rotate the thermostat knob clockwise to increase the temperature counterclockwise to lower the temperature The status lamp is lit when the heaters are energized 2 Allow one...

Страница 10: ...e plug from its outlet for a thorough cleaning between incubation periods The chamber interior and shelves can be cleaned with a mild detergent or soap and water Do not use abrasives or metallic scour...

Страница 11: ...40V units 152 340 209 01 Tension Catch 600 124 00 Hinge 570 341 00 Rubber Feet 4 790 214 00 Shelf 100 105 120 125 Series 172 134 00 Shelf Lower 150 Series 172 135 00 Shelf Upper 120 125 150 Series 172...

Страница 12: ...e products with the replacement parts but only for a period of ninety 90 days from the Commencement Date Lab Line s warranty provided hereunder shall be null and void and without further force or effe...

Страница 13: ...ncubators Ovens Dual Action Open Air Shaker Reciprocating Shakers open air and water bath Rockers and Rotators Low Cost Shakers Environ Blok Shaker Titer Plate Shaker Multi Wrist Shaker Water Baths ex...

Страница 14: ...15 FIRST IN INSTRUMENTS SERVICING SCIENCE INDUSTRY RESEARCH AND EDUCATION SINCE 1908...

Страница 15: ...16 ACCESSORY CHECKLIST The following loose parts and accessories are packed with this unit Before discarding any packing materials please be sure that nothing has been overlooked...

Страница 16: ...100 5JPN 101 105 106 120 120JPN 120 5JPN 121 125 126 172 134 00 _________ Shelf Lower 172 135 00 1 Models 150 150JPN 152 _________ Shelf Upper 172 081 00 1 Models 120 120JPN 120 5JPN 121 125 126 _____...

Отзывы: