7
Maintenance and Service
Page 176
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
NOTICE Observe the following:
Only use cleaning detergents that will not damage the surfaces of
the system.
Never use sharp tools or brushes for cleaning any surfaces.
Do not use sprays for cleaning.
Prevent cleaning detergent from entering the flow path.
Do not use excessively wetted cloth or wipes for cleaning. Prevent
any liquids from entering the functional components of the detector.
Liquids can cause a short circuit when getting in contact with the
electronic components.
Preparations
Turn off the power to the mass spectrometer and disconnect the power
cord from the power source.
Follow these steps
1) Wipe the surfaces clean with a clean, dry, soft, lint-free cloth or wipe. If
necessary, slightly dampen the cloth or wipe with a solution of
lukewarm water and a suitable cleaning detergent.
2) Allow the cleaning detergent to react as recommended by the
manufacturer.
3) Wipe the cleaned surfaces with purified water to ensure that all
cleaning detergent residues have been removed.
4) Wipe the surfaces dry using a soft, lint-free cloth or wipe.
7.4.4
Replacing the ESI Probe Capillaries for the ISQ EC and ISQ EM
Follow the instructions in this section to replace the probe capillaries.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...