6
Operation
Page 164
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
6.8
Remote Start Cable
Remote start cables are used with Hardware Inject Synchronization. This
allows the autosampler or another modules to send a hardware signal
to the MS to tell it to start acquiring at the beginning of a run. Hardware
Inject Synchronization is optional and not used by default.
The MS includes an “LC/IC Start” port that can be used for hardware
synchronization of the run start.
Figure 91: View of the LC/IC Start Port on the MS
The Start Port consists of a 6-pin mini DIN connector. The pins are used as
described in Table 18.
Table 18: Correct use of connection pins
The MS default polarity is to start on an active low signal (pulled to ground to indicate
start of run). In instrument configuration, this can be changed to active high. You must
configure the autosampler or other modules to output the start signal with the same
polarity. Be sure to set the duration of the start signal to 2 s or so.
6.8.1
Vanquish HPLC-Autosampler
Remote start cable part number: 1R76396-1060 (grey cable)
Connect to
MS LC/IC Start port
Pin Number
Used for
1
Not Used
2
Not Used
3
Ground
4
Inhibit Ready Out
5
Ground
6
Remote Start In
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...