6
Operation
Page 162
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
6.7.3
Venting the Mass Spectrometer
Follow the procedure below to vent the mass spectrometer.
1)
Disconnect the data system from the instrument in Chromeleon
software.
2)
Push the power button at the back of the instrument down to the off
position to power off the mass spectrometer. See Figure 89.
Figure 89: Powering off the mass spectrometer
3)
Wait 5 minutes.
4)
Loosen the vent valve knob inside the front door of the instrument next
to the source housing. See Figure 90.