Performing Routine Maintenance
Cleaning the Ion Source Cartridge
68
ISQ 7000 Hardware Manual
Thermo Scientific
Replacing the Ion Source Components
To replace the ion source components and reinsert the ion source cartridge into the
ISQ 7000 mass spectrometer
1. Place the ion volume-repeller insulator on the repeller and hold it in place with the
repeller nut. Set it aside for now.
2. Insert the ion cartridge sleeve into the source holder.
Note
Excessive wear due to cleaning of the repeller can cause the repeller to fit loosely in
the ceramic insulator. Make certain that the repeller is concentric in the insulator before
tightening the repeller nut so that the repeller does not electrically short to the ion
volume.
Figure 59.
Inserting the Sleeve into the Source Holder
Note
If you have an ISQ 7000 mass spectrometer with no VPI, you must order the source
holder separately. See the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering
information.