1
Performing Routine Maintenance
Replacing a Column
16
ISQ 7000 Hardware Manual
Thermo Scientific
Replacing a Column
To replace a column
1. Cool down the GC oven and injector. See the GC documentation for information.
The procedure for replacing a column depends on your system configuration. If your system
does not have a VPI or has an AEI ion source installed, see
7000 system with no VPI or an AEI ion source”
. If your ISQ 7000 MS has a
vacuum probe interlock (VPI) and the ExtractaBrite ion source is installed, see
column on a ISQ 7000 system with a VPI and an ExtractaBrite ion source”
.
Replacing a column in a ISQ 7000 system with no VPI or an AEI ion source
If your ISQ 7000 system has no VPI or has an AEI ion source installed on the system, follow
the instructions below to replace the column.
To replace the factory-installed column in the ISQ 7000 system in a system with no
VPI or with an AEI ion source installed
1. Cool down the GC oven and injector. See the GC documentation for information.
2. Open the ISQ 7000 Dashboard and click
Shut Down
.
During the shutdown procedure the vacuum and heaters lights will remain off. Once the
procedure is complete and the instrument is ready to be powered off, the power light will
turn amber and start blinking rapidly. At this point it is safe to power off the ISQ 7000
system.
Note
If you are running samples, stop the acquisition before powering off the system.
WARNING BURN HAZARD:
The injector, oven, and transfer line may be hot. Allow them
to cool to room temperature before touching them.
ATTENTION RISQUE DE BRÛLURE :
l’injecteur, les détecteurs, le four et la ligne de transfert
peuvent être chauds. Laissez-les refroidir avant de les toucher.
Note
Some of the cleaning procedures in this section require the use of methanol. If
methanol is unavailable or prohibited, substitute LCMS-grade or GC-grade ethanol or
isopropyl alcohol. Do not use denatured ethanol as it may contain impurities that
contaminate the GC-MS system.
Note
If you are running samples, stop the acquisition before powering off the system.