4
Upgrade Equipment
Convectron Gauge and Foreline Adapter with Hose
Thermo Scientific
ISQ 7000 Hardware Manual
395
5. Insert and twist the convectron gauge into the threaded hole on the foreline adapter.
6. Attach the foreline hose on the adapter.
7. Use the rough pump clamp and hose clamp to attach the adapter to the turbo pump.
8. Attach the cable to the distribution board and convectron gauge.
9. Reattach the right side panel and the screws holding it in place.
10. Slide the instrument closer to the GC.
11. Restart your system by following the instructions in
Note
Before you insert the convectron gauge in the hole, wrap Teflon tape around the
part to be inserted. The tape prevents vacuum leaks.
Figure 455.
Installing the Foreline Adapter to the Convectron Gauge
Tape
Convectron Gauge
Foreline Adapter