3
Advanced Maintenance
Replacing the Calibration Gas Components
Thermo Scientific
ISQ 7000 Hardware Manual
363
Calibrant Reservoir Kit
To replace the calibrant reservoir kit, which consists of the calibrant reservoir and its
cover
1. If you are using hydrogen as a carrier gas, unscrew the hydrogen safety screw on the front
door panel.
2. Open the front door panel of the instrument.
3. Twist the calibrant reservoir cover to disconnect it from the calibration gas module.
4. Attach a new calibrant reservoir to the calibration gas controller.
5. Close the front door panel of the instrument.
6. If you are using hydrogen as a carrier gas, replace the front door panel screw.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components. If you just need to refill the calibrant reservoir and not replace it, see
“Refilling the Calibrant Reservoir”
.
Figure 411.
Removing the Calibrant Reservoir and Cover
Figure 412.
Replacing the Calibrant Reservoir and Cover
Calibrant
Reservoir
Calibration
Gas
Calibrant
Reservoir
Cover
Calibrant
Reservoir