Advanced Maintenance
Replacing the Power Supplies
Thermo Scientific
ISQ 7000 Hardware Manual
347
Power Supply System
To replace the power supply system
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Make sure the electrometer board cable is not in the way when you remove the power
supply system.
4. Disconnect the power cables from the back of the mass spectrometer.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 388.
Disconnecting the Power Cables
Power Cable
Connectors
Screws