Powering On the ISQ 7000 Mass Spectrometer
To power-on the ISQ 7000 instrument
1. Install the GC column See
2. Be sure the GC is powered on and there is carrier gas flowing through the column into
the ISQ 7000 instrument.
3. Open the front door of the instrument and tighten the vent valve knob. Be sure not to
pinch the o-ring. Then close the front door of the instrument.
4. Reach around the right side of the instrument and pull up on the power switch to
power-on the ISQ 7000 instrument.
CAUTION INSTRUMENT DAMAGE:
If you power-on the ISQ 7000 instrument without
column flow, air is drawn through the column, which could damage the instrument. This
large air leak may also require that you clean the ion source.
ATTENTION DOMMAGES À L’INSTRUMENT
: si vous alimentez l’instrument ISQ 7000
sans débit de colonne, l’air est aspiré à travers la colonne, ce qui peut endommager
l’instrument. Cette importante fuite d’air peut également nécessiter que vous nettoyiez la
source d’ions.
WARNING FIRE HAZARD:
If you are using hydrogen as a carrier gas, do NOT reach over
the top of the instrument to power it on. Instead, reach around the right side or go to the
back of the instrument and flip up the power switch.
AVERTISSEMENT RISQUE D’INCENDIE
: si vous utilisez de l’hydrogène comme gaz
vecteur, ne passez PAS la main par-dessus l’instrument pour le mettre sous tension. Passez-la
plutôt par le côté droit ou placez-vous derrière l’instrument et activez l’interrupteur
d’alimentation.