Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
336
ISQ 7000 Hardware Manual
Thermo Scientific
Electrometer Board Cable
To replace the electrometer board (15-pin male/female RS-232) cable
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Use a T20 Torxhead screwdriver to remove the two screws holding the support bracket to
the middle of the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 371.
Removing the Controller Interface Board
Support Bracket
Dynode/Multiplier
Power Supply
Support Bracket