Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
334
ISQ 7000 Hardware Manual
Thermo Scientific
Source Interface Board and Heat Shield
To replace the source interface board and heat shield
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. If you are only replacing the heat shield, you do not have to remove the analyzer tray.
Disconnect the filament board and wires from the source interface board and the left
screw to remove the heat shield.
3. Replace the heat shield and reattach the left screw.
4. Reconnect the filament board and wires to the source interface board.
5. If you must replace the source interface board, remove the analyzer tray from the
instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 369.
Replacing the Heat Shield the Filament Board Wires
Note
Keep all manifold parts you remove covered on a clean work surface so they will not
accumulate dust.
Source Interface
Board
Heat Shield
Left
Screw
Filament Board
and Wires
Right
Screw