Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
301
4- Pin Feedthrough
To replace the 4-pin feedthrough:
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. Disconnect the electrometer board cable and large feedthrough cable. Each one is labeled
and permanently connected to the power supply.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 320.
Disconnecting the Electron Multiplier Cables
Electron Multiplier
Plate
Large
Electrometer
Board Cable