Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
299
5. Lift the RF board out of the mass spectrometer. The rod driver board is attached to it.
6. Insert a new RF board/rod driver board kit by aligning the connectors on the bottom of
the board with the connectors on the distribution board beneath it. It will drop into the
slot when it is in the right position.
Figure 317.
Replacing the RF Board/Rod Driver Board Kit
Figure 318.
Reinstalling the RF Board/Rod Driver Board Kit
Connectors