3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
298
ISQ 7000 Hardware Manual
Thermo Scientific
RF Board/Rod Driver Board Kit
To replace the RF board/rod driver board kit
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. The RF board is the second board standing on end toward the center of the ISQ 7000
instrument. Use a T20 Torxhead screwdriver to loosen the screw holding the board and
bracket to the chassis.
4. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering an
RF board/rod driver board kit. This kit must be replaced by a Field Service Engineer, so
when you place an order for the kit, schedule the Field Service Engineer as well.
Note
You do not have to remove the ion source cartridge or the column or vent the
system. If you are suing hydrogen as a carrier gas, power off the GC.
Figure 316.
Removing the RF Board and Bracket
RF Board
Chassis
Screw
Bracket