3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
293
RF Board and Rod Driver Board
To replace the radio frequency (RF) board or rod driver board
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. The RF board is the second board standing on end toward the center of the ISQ 7000
instrument. Use a T20 Torxhead screwdriver to loosen the screw holding the board and
bracket to the chassis.
4. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument.
5. Lift the board out of the ISQ 7000 instrument.
6. Use a T10 Torxhead screwdriver to remove the three screws on the rod driver board. Be
careful not to drop the screws under the metal cover.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering an
RF board or rod driver board. The RF board must be replaced by a Field Service Engineer,
so when you place an order for a new RF board, schedule the Field Service Engineer as
well.
Note
You do not have to remove the ion source cartridge or the column or vent the
system. If you are suing hydrogen as a carrier gas, power off the GC.
Figure 309.
Removing the RF Board and Bracket
RF Board
Chassis
Screw
Bracket
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...