3
Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
278
ISQ 7000 Hardware Manual
Thermo Scientific
Right Side Panel
To replace the right side panel of the ISQ 7000 instrument
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
2. If the ISQ 7000 mass spectrometer is too close to the GC, move them apart so you can
access the right side panel.
3. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge. However, you must remove the
column and vent the instrument.
Figure 292.
Removing the Right Side Panel Screws
Right Side Panel Screws
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...