3
Advanced Maintenance
Replacing Components of the Electron Multiplier
258
ISQ 7000 Hardware Manual
Thermo Scientific
Electron Multiplier Plate
To replace the electron multiplier plate
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Disconnect the electrometer board cable and large feedthrough cable. Each one is labeled.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 270.
Disconnecting the Electron Multiplier Cables
Electron Multiplier
Plate
Large
Electrometer
Board Cable
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...