Advanced Maintenance
Replacing Components of the Manifold
Thermo Scientific
ISQ 7000 Hardware Manual
243
Front Manifold Plate
To replace the front manifold plate
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the vacuum interlock knob and then use a T10 Torxhead screwdriver to remove
the five screws around the knob.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or column. However, you must
vent the instrument.
Note
As you remove the screws around the vacuum interlock knob, watch for a small
spring and safety interlock (location shown below) that will fall out as your remove the
knob. Be sure to catch them and set them aside.
Figure 250.
Removing the Vacuum Interlock Knob
Screws
Vacuum
Interlock Shield
Spring and safety interlock
are behind this screw