3
Advanced Maintenance
Replacing Components of the Ion Source
186
ISQ 7000 Hardware Manual
Thermo Scientific
Ion Source Block
To replace the ion source block
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block, but leave it attached to the tray.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 163.
Disconnecting the Lens Heater Block Grounding Strap
Screws
Lens Heater
Block
Lens Heater Block
Grounding Strap