7.
Close the deck door, then tap Next.
• If you installed a new chip in the sequencer, the sequencer prompts you to
open the sequencing reagents bay doors to empty the waste and remove
used sequencing reagents bay consumables. Proceed to step 8.
• If you are using a chip that was previously installed and has sufficient lane
capacity for the run, the sequencer prompts you to start the run.
IMPORTANT!
The cartridge and bottles in the sequencing reagents bay must be
replaced every time that a new chip is installed, regardless of how many lanes
were used in the previous chip.
8.
Follow on-screen instructions to empty the waste in the Waste Carboy, remove
waste pipette tips, remove the used Genexus
™
Bottle 1, Genexus
™
Bottle 2,
Genexus
™
Bottle 3, and Genexus
™
Cartridge, then tap Next.
IMPORTANT!
Follow all applicable local, state/provincial, and/or national
regulations when recycling or disposing of consumables and liquid waste.
Chapter 7
Load the sequencer and start a run
Load the sequencer and start a run
7
Genexus
™
Integrated Sequencer User Guide