Planned Run workflow key fields
Step or field name
Description
Ion Reporter
If the Ion Reporter
™
Software is installed and enabled, and you want to analyze the
run data using the software, select the
Ion Reporter Account
, then select the Ion
Reporter
™
workflow from the
Existing Workflow
dropdown list.
Research Application
Select the
Research Application
and
Target Technique
that represent your
sequencing experiment.
Instrument
Select
Ion GeneStudio
™
S5 System
.
Chip Type
Select the
Ion 520
™
Chip or Ion 530
™
Chip
.
Library Kit Type
Select the kit used to prepare the library.
Template Kit
Select
Ion 520 & Ion 530 ExT Kit-Chef
.
Sequencing Kit
Select
Ion S5 ExT Sequencing Kit
.
Barcode Set (optional)
If you are using barcodes with:
•
DNA libraries
– Select the appropriate barcode set used.
•
RNA libraries prepared using the Ion Total RNA-Seq Kit v2
– Select the
IonXpressRNA
barcode set, which contains all 16 barcodes in the Ion Xpress
™
RNA
‑
Seq Barcode 1–16 Kit.
If you are
not
using barcodes with:
•
DNA libraries
– Leave the Barcode Set field blank.
•
RNA libraries prepared using the Ion Total RNA-Seq Kit v2
– Select
RNA_Barcode_None
from the dropdown list. This will ensure that the proper
trimming is performed on the resulting sequence when the RNA library does
not have a barcode.
Flows
Enter the appropriate number of flows for the read length (for example, 850 flows
for 400
‑
base
‑
read sequencing and 1,350 flows for 600
‑
base
‑
read sequencing).
Plugins
Select the appropriate plugins for your application.
Projects
Select or add a project to group your run data. You can include runs in multiple
projects, and remove runs from a project at any time.
Run Plan Name
Enter a name for the run.
Reference Library
Select a reference library uploaded to the Torrent Suite
™
Software, if any.
BED files
Select the
Target Regions
or
HotSpot Regions
BED file in the Torrent Suite
™
Software, if any.
Sample Tube Label (Required) Enter or scan the barcode of the Ion Chef
™
Library Sample Tube that you will use to
load your sample into the Ion Chef
™
Instrument.
Chip Barcode
Enter or scan the chip barcode.
Chapter 3
Create a Planned Run
Planned Run workflow key fields
3
22
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide